When the PROFILER™ was introduced
in 1996, it created a new market for real-time beam analyzers with PC
interfaces. The PROFILER can accomplish many of the TG-40 daily, monthly
and annual tests more efficiently than any alternative technique. The
PROFILER™ has established itself as a gold standard in the QA market due
to the fact that it is easier to use, and provides more data at a lower
price than competing products. PROFILER™ applications are as specific or
diverse as the user chooses. Routine QA issues, such as those found in
TG-40, are quickly addressed with the PROFILER's 46 channel real-time
linear array
|
|
|
| |
|
Physical Description
The PROFILER™ is a linear array of 46 radiation hardened diode detectors.
The detectors are spaced at every 5 mm, and are suitable for both Photon and
Electron energies. Beam size at 100SSD is 20cm x 20cm.
The PROFILER™ diode array is imbedded beneath 1 cm of intrinsic H2O
equivalent buildup. The entire product weighs only 8.5 pounds, and is used
directly on the treatment couch, or in the available gantry attachment.
Applications
PROFILER™ applications range from Dynamic or Virtual Wedge commissioning and
Tissue Compensator QA, all the way to Monthly Flatness checks and beam
troubleshooting.
-
TG-40 QA (Flatness, Penumbra, Area Symmetry, Mirror Symmetry,
Light/Radiation Field Coincidence, Field Size, Beam Center)
-
IMRT QA (Snap-Line Utility compares Profile measurement to treatment plan
output, H & D curve calculation for Radiological Imaging Technology 113 when used with Gantry Phantom)
-
Beam Troubleshooting & Beam Steering Adjustments
-
VW & EDW QA
-
Tissue Compensator QA
PC Communication
The PROFILER™ communicates with a host computer via a single 25 M cable. The
host computer is the comprehensive interface and storage device for the
PROFILER™, which supports Windows 95 and higher, as well as Windows NT.
Real-Time Analysis
PROFILER™ data is viewed on the PC screen in real time, and can be viewed in
either dose or rate mode. Profiles just taken can be easily compared against
earlier profiles with the click of a mouse. Automatic analysis routines
provide a wealth of data viewable in real time as the beam is on.
Data Storage
Profiles are stored on PC.
Accuracy
The PROFILER™ uses pre-irradiated p-type diode detectors that offer
exceptional long-term stability. For 6MV beams, diode sensitivity will
change less than 2% over 100,000 MU.
Accessories
The Profiler is available with many useful accessories that will enhance
it's functionality.
-
MDA - Motor Drive Accessory for remote
positioning
-
TP & Compensator Verification Utility
-
Import Treatment Plans and Tissue Compensators.
-
Gantry Attachment
-
Electron Energy Wedge
Examples of Use
Features
Instant Real-Time data for fast data acquisition, analysis, and
verification. All-in-one design contains detectors and electronics – easy
setup.
Single power/data cable is the only connection required.
Available Gantry Attachment for measurements at any gantry angle.
Available Motor Drive for automatic remote positioning and 2-D graphing.
Lightweight – only 8.5 pounds.
|