Q: What are the main differences between my
existing Profiler and the new PROFILER 2?
A: The PROFILER 2 features some significant enhancements from the
widely acclaimed Profiler. Some key changes include:
- The addition of an X-axis - the Profiler 2 now has a Y-axis with 83
detectors and an X-axis with 57 detectors
- Higher resolution – the spacing between detectors has been reduced from
5mm to 4mm
- Larger field size – at 100 SSD the maximum field size is 32.8 x 22.4cm
- Inherent buildup – the detector array is imbedded beneath 1cm of
Virtual Water™
- New software – the new PROFILER 2 comes with new software which offers
more versatility and ease-of-use
Q: Is there a gantry fixture I can use with the
PROFILER 2 to measure at any gantry angle?
A: Yes. The PROFILER 2 has been designed so that it fits into the IMF (Isocentric Mounting Fixture) which can also be used with the MapCHECK
and the Daily QA 3. The IMF holds the Profiler 2 at 100cm SDD and
accommodates enough buildup to measure at 10cm depth. Rotational
measurements can be done at any gantry angle while keeping the PROFILER 2
securely fixed at isocenter. This is a fast and precise routine for MLC,
Collimator, and Gantry shift effects.
Q: Can I measure a larger field such as 40cm using
the PROFILER 2?
A: Yes. Larger field sizes are possible by taking two
measurements at 100cm SSD using the concatenate feature in the software, or
a single measurement can be taken at a shorter SSD.
Q: Does the PROFILER 2 use ion chambers or diode
detectors?
A: The PROFILER 2 uses diode detectors measuring only 0.64mm˛.
The small size ensures accurate measurements which is particularly critical
in the penumbra region. The detectors have the highest available sensitivity
and lowest available volume of any electronic measurement system.
Q: How frequently do I have to calibrate the
PROFILER 2 detector array?
A: As there is very little change in the sensitivity of the
detectors (±0.5% per 100,000 cGy on a 6 MV beam), the PROFILER 2 requires no
frequent calibration. The Profiler is calibrated using a patented wide field
calibration routine, which assigns each detector a unique sensitivity
correction factor. This procedure takes approximately 15 minutes and is
typical performed once a year. The calibration is done by the user by
following simple step by step instructions in the PROFILER 2 software.
Q: Can I import plans from my TPS to make
comparisons?
A: Yes. The PROFILER 2 will import the treatment planning file
from a TPS and perform comparisons of the planned and measured dose. A
percent difference can be instantly viewed by selecting both the measured
and planned file and then clicking on compare. The PROFILER 2 supports all
major TPS formats.
Q: Can I use my PROFILER 2 for electron energy
verification?
A: Yes. By using the optional aluminum Electron Energy Wedge you
can quickly and accurately verify electron energy. The PROFILER 2 software
calculates the electron energy by slope or intercept analysis.
Q: How can I use the PROFILER 2 to study the
characteristics of transient events during warm-up?
A: The PROFILER 2 can take measurements using frame mode which
permits profile measurements at rates from 1 to 15 measurements per second.
The playback options let you pay back the collected profiles at variable
speeds, using the frame capture playback bar in the software. This feature
may be particularly important in IMRT applications.
Q: I have a MapCHECK, can I use the same power/data
cable for the PROFILER 2?
A: Yes. MapCHECK, TomoDose, Profiler, PROFILER 2, Daily QA Check
2 and Daily QA3 all use the same 25 meter power/data cable which
enables the user to simply switch the instruments. This keeps the number of
cables to a minimum when using SNC products.
Q: Can I use the PROFILER 2 for Daily QA?
A: The PROFILER 2 will have the ability to take Daily QA
measurements, but will not have a daily QA database built into the software.
However, the new PROFILER 2 software does display interpolated Daily QA3
measurement data from the selected data file. This Daily QA3 data can also
be printed. For Daily QA we always encourage our users to use the Daily QA
Check 2 or the soon to be launched Daily QA3, which are specifically
designed for morning QA. The Daily QA Check 2 and Daily QA3 use
ion chambers which SNC use for products that receive high levels of
radiation daily, and where precise penumbra measurement is not needed. The
PROFILER 2 uses diode detectors
because of their high precision and accuracy for the measurement of profiles
and steep dose gradients.