Q: What is the diode cable length?
A: The standard cable length is 3 Meters. If a longer cable is
desired, 10 Meter extension cables are available.
Q: What is the expected lifetime of the diode?
A: As long as the diode and cable are treated with care, lifetime
should be at least 10 years or 1,000,000 cGy. The most common cause of diode
failure is cable or connector failure.
Q: Do I need negative or positive output diodes?
A: The diode output should match the input of your electrometer. Sun
Nuclear’s rf-IVD™ and Model 1137 IVD™ are polarity independent.
Q: Will the diodes work on my electrometer that was
made by a different manufacturer?
A: Yes.
Q: Why does the diode include buildup?
A: This is provided as a convenience. Most patient dose measurements
are desired at dmax. The buildup for each diode is chosen to return a result
that is close to dmax for the particular energy range of the diode used.
Without standard buildup, the measured data would have to be manually
corrected for dmax. If for some reason this approach is desired, Sun Nuclear
does manufacture a diode without buildup – The QED™ Skin Detector.
Q: What is the sensitivity of the diode?
A: 0.4nC/cGy
Q: What is the reproducibility of the diode?
A: 99.5% reproducible for >1 cGy
Q: Which diode style does Sun Nuclear sell more of:
QED or ISORAD-3™?
A: Overall sales between the QED and ISORAD-3™ are evenly split.
Q: How do the Sun Nuclear diodes compare to the
MOSFET system?
A: Follow this link to learn more:
MOSFET vs DIODES