How EPIDose Works
EPIDose introduces an EPID-to-Dose conversion that allows your IMRT QA to remain dose-based, resulting in accurate conversions of EPID images to "Dose Planes". EPIDose operating in the MapCHECK analysis environment, offers complete independence from the delivery and planning systems along with a suite of familiar and proven analysis options.
EPID Based QA with EPIDose
- Using a raw EPID file, the EPID image is projected to the desired dose plane location. Corrections for output factors and the source distribution of scattered photons are applied to each MLC sub-field.
- The results are convolved with a dose redistribution kernel which converts the dose from EPID response (sharper than water) to a water equivalent scattering of dose.
- The results are then automatically calibrated to dose using data in the EPIDose model (based on MapCHECK absolute dose planes).

Physics Modeling
Physics Modeling is quickly performed for each linac and energy combination. After EPIDose physics modeling is initially done, it is not required again unless there is a change in the EPID.
Product Options


