Sun Nuclear - Your Most Valuable QA and Dosimetry Tools
 


Quick Links

 

Resources
 

Related Products
 

 
  • Home
  • Specifications
  • Software
  • FAQ

The PC Electrometer Advantage
  • World’s smallest reference class electrometer
    • Compact size, lightweight - 0.4kg
    • Industry leading specifications
    • High performance - Fast sampling - 100 μs
  • USB powered - no batteries or power cord
  • Quick to setup and easy to use
  • Fully configurable software interface
  • Interfaces with the Sun Nuclear 1D Scanner™
Hardware
  • Two independent measurement channels
  • Less than one minute warm up time
  • Temperature probe input
  • Single cable connection options
    • USB or Sun Nuclear standard power/data cable
PC Electrometer sampling interval is 100 µs. Each step in this graph of total charge indicates a linac pulse as measured with a 0.6cc chamber. PC Electrometer clearly shows the increase in linac pulse frequency as a function of dose rate. From 80 MU/min, the additional pulses are 1, 2, 3, and 4 for the respective rates of 160, 240, 320, 400.
Fast sampling makes PDD curves smooth and accurate when measured with the PC Electrometer and 1D Scanner. The PDD curve below is an un-smoothed ratio of Field to Reference, using 0.6cc Farmer type chambers.

Features and Specifications

Software
  • Unique configurable design
  • Auto Start/Stop feature that starts data acquisition
    automatically when the beam is detected, and stop
    data acquisition automatically when the beam turns off
  • Depth Scan functionality with the 1D SCANNER
  • Hide/show different functions
  • Automatic beam type sensing
  • Timed measurement option
  • Expand data display for distance viewing
  • Detector library stores calibration factors

Frequently Asked Questions

Q: Is the PC Electrometer a reference class electrometer?

A: Yes. The PC Electrometer was tested for compliance with the applicable performance specifications documented in the IEC 60731 standard. Click here for a table showing the PC Electrometer specifications and the IEC 60731 reference class standard specifications.


Home  |  Patient  |  Machine  |  Routine  |  Dosimetry  |  Radon  |  Support  |  Education  |  Downloads  |  Company  |  Contact Us